Enterprise Qualifications & Standards

Hebei Huanqing Monitoring Technology & Instrument Center strictly follows national environmental industry standards and has established enterprise standards to ensure product quality and performance meet CEMS system requirements.

Enterprise Standard

Q/HBHQ 01—2024 “Portable Pre-treatment Equipment”

Specifies terminology, product classification, technical requirements, test methods, inspection rules, marking, packaging, transportation, and storage for portable flue gas pre-treatment equipment.

Core Technical Specifications

Key technical parameters of Huanqing portable pre-treatment equipment per enterprise standards:

Parameter Specification
Temperature Control Accuracy ±0.2℃
Sample Gas Output Dew Point ≤4℃
Probe Inlet Temperature ≤200℃ / ≤1000℃ / ≤1800℃ (varies by model)
Filtration Precision 0.3μm / 1.2μm / 5μm / 10μm (varies by filter)
Flow Range 0.5~6 L/min
Rated Power 400W / 650W / 800W (varies by model)
Protection Rating IP20
Operating Voltage AC220V±10% 50/60Hz
Ambient Temperature +5℃~+43℃

Quality Assurance

All products undergo rigorous type testing and factory inspection before delivery, including:

  • Appearance inspection: smooth surface, no burrs, no deformation, clear labeling
  • Insulation resistance: ≥2MΩ
  • Dielectric strength: AC1500V without breakdown or flashover
  • Temperature control accuracy: ±0.2℃
  • Sample gas output dew point: ≤4℃
  • Sealing: no leakage
  • Flow indication error: within ±5%
  • Filtration precision & efficiency: meets corresponding filter grade requirements

Packaging & Transportation

Products are packaged with shockproof and moisture-proof materials. Package labels include product name, model, quantity, and manufacturer. Avoid severe vibration and collision during transportation. Storage conditions: well-ventilated, dry, temperature -10℃~+55℃, relative humidity ≤85%.